ba0006p089 | (1) | ICCBH2017
Mandlik Rubina
, Kirtikar Utkarshini
, Ekbote Veena
, Jaiswal Anjali
, Kadam Nidhi
, Khadilkar Vaman
, Chiplonkar Shashi
, Khadilkar Anuradha
Objective: The objective of this study was to enable migration of DXA data of patients following replacement of the GE Lunar DPX-Pro, pencil beam densitometer with GE Lunar iDXA, fan beam densitometer.Methods: Scans of total body were conducted for 67 subjects, aged 3.0 to 19.9 years, on both machines. A sub-set of six subjects (3 boys) were randomly selected and reserved as a validation sample. Differences between the values obtained on the two machines...